Management and microbial monitoring of final rinse water for flexible endoscopes in 290 hospitals in Jiangsu Province, China: a multicenter cross-sectional study

Abstract

Background

There is a lack of a universally accepted standard or guideline for the frequency of disinfection in purified water pipelines. Furthermore, there is no standardized method for detecting microorganisms in the final rinse water utilized for endoscope cleaning.This study aims to examine the current management and microbial monitoring practices concerning the final rinse water used for flexible endoscope cleaning in medical institutions.

Methods

A questionnaire was designed using a convenience sampling method to gather data on the maintenance and microbial monitoring of final rinse water for flexible endoscopes in 290 medical institutions across Jiangsu Province, China.

Results

Purified water is used for endoscope rinsing by 93.45% of institutions, with 78.62% employing centralized water supply. Membrane filtration devices at the terminal are installed by 82.07%, mainly with a 0.2μm pore size (76.47%), and are replaced quarterly (32.77%). Disinfection devices are present at 52.76% of terminals, with varied disinfection frequencies; chlorine-containing disinfectants (48.15%) and peracetic acid (34.92%) are predominant. Inadequate disinfection, filter membrane neglect, sampling contamination, and biofilm formation are identified as reasons for non-compliant final rinse water. Actions include filter replacement, pipeline disinfection, and flushing. Microbial mornitoring occurs quarterly (70.96%), with faucet outlets as primary sampling sites. Standards are based on 10cfu/100ml (87.58%), using membrane filtration (40.81%) and nutrient agar plates (82.72%). A cultivation period of 2 days predominated (72.43%), with a temperature range of 35-37°C (76.47%).

Conclusion

While purified water and terminal filters are common for final rinsing of endoscopes, there is variation in maintenance and supply line disinfection. Current microbiological methods’ reliability is considered low, necessitating further research to establish unified standards for effective endoscope final rinse water management and monitoring.

Introduction

With continuous updating and upgrading, flexible endoscopes have integrated examination and operation functions [1]. More and more complex features bring complexity to the structure, which in turn increases the difficulty of reprocessing [2] The risk of endoscopic-related infection has been listed as one of the top ten health technology hazards released by ECRI (Emergency Care Research Institute) Institute in 2018 and 2019 [3]. Endoscope-related infections are associated with factors such as failures in the reprocessing of flexible endoscopes, equipment-related issues (including endoscope automated cleaning and disinfection machines, pure water systems, and storage equipment), product malfunctions, and human errors [4]. At the same time, microbial contamination of final rinse water has also been associated with outbreaks [5] and pseudo-outbreaks [6,7,8,9,10], involving some flexible endoscopes.

The main purpose of final rinsing is to remove the residual chemical disinfectant to avoid adverse reactions to patients or workers. However, when the final rinse water is contaminated with bacteria or other pathogens, this step risks re-contaminating the endoscope and possibly causing infection to the patient. A multicenter study led by Zhongshan Hospital affiliated to Fudan University [11] and Tianjin Center for Disease Control and Prevention [3] (encompassing medical institutions at various levels) revealed that the rates for qualified final rinse waterr (water that meets the expected quality for the final rinse of endoscope reprocessing) were 63.09% (53/84) and 61.11% (111/180) respectively. The highest colony-forming units (cfu) per 100 mL of a single specimen reached as high as 91,000 cfu/100 mL. These results indicate it is essential to implement microbiological controls to ensure the final rinse water remains high quality.

There is currently no uniform standard for the maintenance requirements of the final rinse water processing system in China, and the microbial monitoring methods of the final rinse water are not consistent across medical institutions. The aim of this study was to investigate medical institutions in the Jiangsu Province and examine the water management and microbial monitoring for final rinsing of flexible endoscopes, to identify the existing issues and provide a basis for unifying requirements and establishing standards.

Methods

Objective

This study included a questionnaire developed to collect information on the management and microbial monitoring of final rinse water used for reprocessing flexible endoscopes in medical institutions. A total of 290 medical institutions in Jiangsu Province, China were included in this study. Out of these, 129 (44.48%) were tertiary hospitals and 161 (55.52%) were secondary hospitals. These i nstitutions were spread across 13 prefecture-level cities in Jiangsu Province. Specifically, there were 53 institutions in Nantong, 45 in Suzhou, 35 in Yancheng City, 28 in Huaian City, 21 in Taizhou City, and 18 in Nanjing and Changzhou each. Additionally, there were 15 institutions in Zhenjiang, 14 in Suqian, Yangzhou, and Wuxi, respectively, 8 in Xuzhou, and 7 in Lianyungang (Fig. 1).

Fig. 1
figure 1

Mapping representation of study area. A The location of Jiangsu Province in China. B Distribution of Hospitals Participating in This Study Across Jiangsu Province

Questionnaire design

To resolve practical difficulties arising in field operations, the questionnaire was designed based on relevant regulations and literature reports. It was refined through review feedback from two experts from endoscopy centers and four senior infection control specialists. The final version of the questionnaire comprised sections on: (1) the information about the final rinse water for flexible endoscopes; (2) configuration and maintenance of reprocessing processing equipment; (3) analysis of reasons for non-compliance and corresponding responses; (4) and the method of microbial monitoring for flexible endoscopes and final rinse water. See the supplementary table for details.

Sampling method

The questionnaire was created using the ‘Questionnaire Star’ platform and a Quick Response(QR) code was generated. It was published to the WeChat Group of the Fifth Hospital Infection Management Professional Committee of Jiangsu Hospital Association on June 26, 2023. Participation was requested from secondary and higher-level medical institutions in the province, with each medical institution required to complete only one questionnaire and assign personnel for its completion. A convenience sampling method was employed, where all institutions in the WeChat group were invited to participate. This method was chosen for its practicality and efficiency in reaching a broad range of institutions within a limited timeframe. A total of 302 questionnaires were collected and underwent a rigorous data cleaning process. Two individuals independently reviewed each questionnaire, and telephone confirmation was conducted when necessary. During this process, ten duplicate questionnaires and two questionnaires with missing data were excluded, resulting in a final set of 290 valid questionnaires for analysis.

Statistical analysis

Statistical analysis was conducted using the WPS2019 (Kingsun. Hong Kong, CHINA) for data processing and the SPSS20.0 (IBM Corp. Armonk, NY,USA) for data analysis. Counting data were presented as frequency and percentage, and inter-group comparison was performed using the Chi-square test. A statistically significant difference was considered when P < 0.05.

Results

Current status of flexible endoscope and final rinse water

In 290 medical institutions, the types of flexible endoscopes used included gastroscopes (99.31%) and colonoscopes (96.21%), which were equipped in more than 95% of institutions, while fiberoptic bronchoscopes (56.21%) and laryngoscopes (51.03%) were equipped in more than 50% of institutions.” Among them, the proportion of duodenoscopes, fiberoptic bronchoscopes, laryngoscopes and esophageal ultrasound endoscopes in tertiary hospitals was significantly higher than that in secondary hospitals. As to manual cleaning and disinfection workstations for flexible endoscopes, tertiary hospitals mostly had 3 to 5 stations (59.69%), while secondary hospitals mostly had fewer than 3 stations (75.78%). In tertiary hospitals, the majority (55.81%) had 1–5 units of automated cleaning and disinfection machines for endoscope reprocessing. The majority of secondary hospitals had no endoscope machines. Manual cleaning and disinfection workstations were mainly located in endoscope centers (97.59%), Ear, Nose, and Throat(ENT) outpatient clinics (30.00%), department of pulmonary and critical care medicine (13.10%) and operating rooms (13.10%). The proportion of purified water (including fully purified water and partially purified water) for final rinsing of flexible endoscope was 93.45%, and the proportion of centralized water supply (including fully centralized water supply and partially centralized water supply) was 78.62%. 78.97% of medical institutions were equipped with water storage tanks for final rinse water (including full configuration and part configuration). See Table 1 for details.

Table 1 Current status of flexible endoscope and final rinse water in 290 medical institutions

Configuration and maintenance on terminal equipment of final rinse water for flexible endoscope

The installation of membrane filters at the terminal of water for final rinse is a critical step in ensuring the microbial quality of the rinse water. These filters are placed at the point where the water enters the faucet (before water discharge), reducing the risk of contamination during the endoscope reprocessing process. 82.07% of medical institutions installed membrane filters at the terminal of water for final rinse (before water discharge), with the main pore size of 0.2μm (76.47%). The main replacement cycle of the membrane filters was quarterly (32.77%), no fixed cycle (26.05%), yearly (21.43%) and semi-annually (14.29%), mainly by third-party companies (51.06%) and hospital equipment/general affairs/logistics. 52.76% of medical institutions installed disinfection devices at the terminal of water for final rinse (before water discharge), with the main disinfection methods including chemical disinfection, ultraviolet disinfection, and ozone disinfection. As to the water supply pipeline disinfection frequency, 34.83% did not disinfect, 18.62% disinfected quarterly, 17.59% had no fixed disinfection frequency, and the main disinfectants were chlorine-containing disinfectants (48.15%) and peracetic acid (34.92%). See Table 2 for details.

Table 2 Configuration and maintenance of terminal equipment on final rinse water for flexible endoscope in 290 medical institutions

Analysis of the reasons for unqualified water used for the final rinsing of flexible endoscopes and corresponding measures

According to the relevant guidelines [12], the qualified standard for microbial culture results of final rinse water for flexible endoscopes is≤10cfu/100 ml. Results exceeding this threshold are considered unqualified. Among the 290 medical institutions, 18 did not detect the final rinse water for flexible endoscope, and 3 did not have the unqualified water for final rinsing. Excluding these, a total of 269 medical institutions were included in the partial analysis, including 125 tertiary hospitals and 144 secondary hospitals (Table 3). In the analysis of the reasons for the unqualified results of final rinse water for endoscopes, the possible factors included ineffective disinfection of the water supply pipeline, no replacement of terminal filter membrane, sampling contamination and biofilm formation at the faucet end of water supply pipeline. It is important to note that these factors are based on institutional analysis and perception, and definitive evidence, such as laboratory testing for biofilm formation or other confirmatory measures, was not consistently available across all cases. Replacement of terminal filter membrane, disinfection of water supply pipeline, and flushing of water supply pipeline were the main measures taken when the water used for final rinsing was deemed unqualified. See Table 3 for details.

Table 3 Analysis of the reasons for unqualified water used for the final rinsing of flexible endoscopes and corresponding measures in 290 medical institutions

Current situation of microbial detection methods for flexible endoscope and final rinse water

Only 3.45% of medical institutions had independent laboratories for testing final rinse water, and 27.24% of medical institutions outsource the testing of endoscope and terminal rinse water samples to third-party companies; Infection control specialists were responsible for testing and cultivating throughout the entire process in only 14.14% of medical institutions. The detection frequency of final rinse water was mainly once every quarter (70.96%), the outlet of faucet and irrigation port accounted for 90.81% and 28.68% of the sampling sites, and the largest sampling quantities were 10 ml (38.6%), 50 ml (33.46%) and 100 ml (14.14%), respectively. The standard of eligibility was primarily based on 10cfu/100 ml (87.58%). The membrane filtration method accounted for 40.81% of the inoculation methods. A cultivation period of 2 days predominated (72.43%), with a temperature range of 35–37°C (76.47%). The selection of culture dishes primarily favored nutrient agar plates (82.72%) See Table 4 for details.

Table 4 Current situation of detection methods on final rinse water of flexible endoscope in 290 medical institutions

Discussion

The WS 507–2016 Technical regulations for cleaning and disinfection of flexible endoscopes [12] (referred to as technical specifications) defines the final rinsing as the process of rinsing the endoscope with purified water. Research findings indicate that 93.45% of flexible endoscopes are rinsed with purified water (fully or partially), with tertiary hospitals using purified water in 98.45% of cases and secondary hospitals in 89.44% of cases. This showed that the majority of medical institutions surveyed in the Jiangsu Province are using purified water as the final rinse water, which was higher than the results of a study conducted by Zhu Xuanrui et al. [13] in 2020 in 72 medical institutions in Jilin Province (87.9% for tertiary hospitals and 80.8% for secondary hospitals), as well as than the results of a study conducted by Liu Feng et al. [14] in 2020 in 318 medical institutions in Shandong Province (53.88% for tertiary hospitals and 30.84% for secondary hospitals). Furthmore, this study found that the frequency of testing for final rinse water was primarily quarterly (70.96%) and monthly (14.34%), which is higher than the results of a survey [15] conducted in 165 medical institutions across 39 countries, where the primary frequencies were monthly (22%) and semi-annually (15%). At present, there is no clear requirement for the sampling frequency of final rinsing water, and medical institutions may set it according to their specific circumstances.

While purified water equipment is highly effective in removing microorganisms and impurities during the initial purification process, the quality of purified water can be compromised as it travels through extensive pipelines to reach the end-use point. Potential contamination risks include biofilm formation, pipeline degradation, and prolonged water stagnation within the pipes. These factors may contribute to microbial overgrowth, ultimately undermining the integrity of the purified water. To address this, additional filtration at the end of the final rinse water and the maintenance of the purified water pipeline are common and necessary practices. These ensure that the water used for the final rinse meets the required microbial and chemical standards, even after potential contamination during distribution. This step is critical for maintaining the safety and efficacy of endoscope reprocessing [12].

In this study, it was found that 82.07% of the final rinse water outlets for flexible endoscope rinsing were equipped with a filter membrane..Among these, 84.50% were in tertiary hospitals and 80.12% were in secondary hospitals, which were higher than the results reported by Zhu Xuanrui et al. [13] (81.80% in tertiary hospitals and 53.80% in secondary hospitals).According to the technical specifications [12], the pore size of the filter membrane used in the production of purified water should be ≤ 0.2μm and should be replaced regularly.The research discovered that 93.28% of hospitals complied with the technical specifications, with 96.33% of tertiary hospitals and 90.70% of secondary hospitals meeting the requirements. Most of medical institutions could meet technical requirements for membrane pore size. It is important to note that the filter membrane traps bacteria instead of killing them, and therefore, it needs to be regularly replaced to maintain its filtering effect. If it is changed too infrequently or never changed, filter membrane will become a source of contamination [16, 17]. In this research, only 1.26% of medical institutions never replace their filters. The majority of medical institutions had replacement frequencies concentrated on a quarterly basis (32.77%), no fixed cycle (26.05%), and yearly (21.43%). This replacement cycle is lower than the findings [13] of Zhu Xuanrui et al. The replacement frequency of the filter membrane in various medical institutions needs to be integrated with the requirements of the manual, water consumption, water quality and the monitoring results of the final rinsing water.

The maintenance of the purified water pipeline is crucial for controlling microbial indicators. However, there is currently a lack of multicenter studies on the disinfection interval of the purified water pipeline, and there is no unified standard or guideline in China [18]. Some studies [19, 20] recommend monthly disinfection of the water supply pipeline system to ensure the quality of purified water. However, further research and discussion are needed to determine its applicability, considering various factors such as raw water quality, temperature, and environmental conditions. This study found that the disinfection frequency of water supply pipelines varied, with some medical institutions indicating no disinfection was being performed..Among those medical institutions that do perform disinfection, the frequency was inconsistent indicating a lack of clear guidance for each medical institution. Therefore, it is crucial to provide relevant evidence-based research and standards to guide them in this matter.

Studies have shown that the primary causes of contamination in final rinse water is the failure to regularly replace the filter [7, 8, 21, 22], improper filter pore size [23], filter missing [24] and ineffective disinfection of water supply pipelines [25]. This is consistent with the findings of this survey. In this study, the main measures taken to rectify poor testing results were disinfection of the water supply pipeline and replacement of the terminal filter membrane. Furthermore, it has been reported that continuous hypochlorous acid disinfection [26] and the installation of an ozone and ultraviolet dual system [27] can significantly enhance the qualified rate of microbial detection in the final rinse water by not only producing qualified final rinsing water but also disinfecting the water pipeline. Therefore, these factors and measures can serve as valuable references when dealing with contamination in the final rinse water.

The filtration membrane method is recommended for the detection of pure water according to relevant regulations both domestically [28] and internationally [29]. This method is highly effective due to its sensitivity, capability to process large sample volumes (which increases the likelihood of detecting low bacterial concentrations), and its ability to reduce interfering substances (e.g., disinfectant residues). By concentrating bacteria on a membrane, it allows for more reliable detection of low bacterial concentrations compared to traditional methods. Its direct culturing approach minimizes errors, and its standardized procedure ensures consistent results, making it an ideal choice for monitoring water quality in medical settings. In this study, only 40.81% of medical institutions utilized the filtration membrane method for the microbial detection of endoscopic final rinsing water. This finding underscores the necessity for further standardization and promotion of the filter membrane method. Additionally, regarding the sample volume for the final rinse water, domestic guidelines do not provide clear recommendations. However, the relevant British guidelines [29] recommend a volume of 100 ml, as a larger sample size increases the likelihood of detecting microorganisms in the water. This offers a valuable reference for further standardizing the sample volume selection when using the filtration membrane method for sampling.

Currently, there is no standardized culture method for detecting microbes in the final rinse water for endoscopes (including culture temperature, culture time, and plate type). There are three main methods commonly used in the laboratory for culturing. The first method is based on the culture method of drinking water [30], which involves culturing the water at 35–37°C for 48 hours on a nutrient agar plate [31, 32]. The second method is based on the culture method of dialysis water (YY0572) [33], where the water is cultured at 18–23°C for 7 days on low-nutrient agar plates such as R2 A [34, 35]; The third method refers to the culture method described in the Chinese Pharmacopoeia for purified water [28], which involves culturing the water at 28–32°C for 5 days on low-nutrient agar plates like R2 A [11, 18]. The British guideline [29] also recommends culturing the water at 28–32°C for 5 days on low-nutrient agar plates.

Culturing purified water using a low-nutrient medium, lower temperature, and longer duration is more suitable as excessive nutrients can inhibit the growth of heterotrophic bacteria and reduce the detection rate [36]. Previous studies [34] have shown that culturing on low-nutrient agar plates at 20℃ for 7 days yields a significantly higher microbial detection rate compared to culturing on normal nutrient agar at 36℃ for 48 hours. However, there is currently no comparison between the second and third methods, which is an area that requires further investigation. This research has identified significant differences in the microbial detection methods used for the final rinse water of endoscopes. The membrane filtration method accounted for less than half of the inoculation methods, with a culture time of mainly 2 days and a culture temperature of mainly 35–37℃. Nutrient agar plates were predominantly selected for culturing These findings indicate that approximately 80% of medical institutions are potentially using unreliable detection methods for the final rinse water, highlighting the need for verification and improvement of these methods.

Conclusions

This research discovered that the ratio of purified water and terminal filter membrane configuration is relatively high for the final rinse water of flexible endoscopes. However, there are notable variations in the frequency of filter membrane replacement and disinfection of the water supply pipeline. The reliability of the current microbial detection method is limited, underscoring the necessity for further research to establish standardized requirements and protocols. This will enhance the effectiveness of monitoring the final rinse water used in endoscope reprocessing.

Data availability

No datasets were generated or analysed during the current study.

References

  1. Schulman AR, Howell JD. From Hirschowitz to 2023: modern endoscopy and beyond. Clin Gastroenterol Hepatol. 2023;22:684–8.

    Article  PubMed  Google Scholar 

  2. Cattoir L, Vanzieleghem T, Florin L, Helleputte T, De Vos M, Verhasselt B, et al. Surveillance of endoscopes: comparison of different sampling techniques. Infect Control Hosp Epidemiol. 2017;38:1062–9.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Ji XY, Ning PY, Fei C-N, Liu J, Liu H, Song J. The importance of sampling technique and rinse water for assessing flexible gastrointestinal endoscope reprocessing: a 3-year study covering 59 centers. Am J Infect Control. 2020;48:19–25.

    Article  PubMed  Google Scholar 

  4. Calderwood AH, Day LW, Muthusamy VR, Collins J, Hambrick RD, Brock AS, et al. ASGE guideline for infection control during GI endoscopy. Gastrointest Endosc. 2018;87:1167–79.

    Article  PubMed  Google Scholar 

  5. Kovaleva J, Peters FTM, van der Mei HC, Degener JE. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Rev. 2013;26:231–54.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Mitchell DH, Hicks LJ, Chiew R, Montanaro JC, Chen SC. Pseudoepidemic of Legionella pneumophila serogroup 6 associated with contaminated bronchoscopes. J Hosp Infect. 1997;37:19–23.

    Article  CAS  PubMed  Google Scholar 

  7. Rossetti R, Lencioni P, Innocenti F, Tortoli E. Pseudoepidemic from Mycobacterium gordonae due to a contaminated automatic bronchoscope washing machine. Am J Infect Control. 2002;30:196–7.

    Article  PubMed  Google Scholar 

  8. Seidelman JL, Wallace RJ, Iakhiaeva E, Vasireddy R, Brown-Elliott BA, McKnight C, et al. Mycobacterium avium pseudo-outbreak associated with an outpatient bronchoscopy clinic: lessons for reprocessing. Infect Control Hosp Epidemiol. 2018;40:106–8.

    Article  PubMed  Google Scholar 

  9. Bennett SN, Peterson DE, Johnson DR, Hall WN, Robinson-Dunn B, Dietrich S. Bronchoscopy-associated mycobacterium xenopi pseudoinfections. Am J Respir Crit Care Med. 1994;150:245–50.

    Article  CAS  PubMed  Google Scholar 

  10. Zhang Y, Zhou H, Jiang Q, Wang Q, Li S, Huang Y. Bronchoscope-related Pseudomonas aeruginosa pseudo-outbreak attributed to contaminated rinse water. Am J Infect Control. 2020;48:26–32.

    Article  CAS  PubMed  Google Scholar 

  11. Shi Q, Hu B, Cui Y, Sun W, Shen Y, Chen X, et al. Investigation on the current situation of terminal rinsing water for flexible endoscopes in 30 tertiary medical institutions in Shanghai. Chin J Nosocomiol. 2020;30:923–6.

    Google Scholar 

  12. National Health and Family Planning Commission of the People's Republic of China. Technical regulations for cleaning and disinfection of flexible endoscopes (WS507-2016). Beijing; 2016.

  13. Zhu X, Zhuang Y, Liu W, Wang X. Investigation on the current situation of water management for terminal rinsing of flexible endoscopes in 72 medical institutions in Jilin Province. Chin J Infect Control. 2022;21:642–50.

    Google Scholar 

  14. Liu F, Li M, Yuan Y, Lian J, Yin J. Investigation on the current situation of water usage for flexible endoscopes in 313 medical institutions. Chin J Disinfect. 2021;38:678–80.

    Google Scholar 

  15. Kenters N, Tartari E, Hopman J, El-Sokkary RH, Nagao M, Marimuthu K, et al. Worldwide practices on flexible endoscope reprocessing. Antimicrob Resist Infect Control. 2018;7:153.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Wang C, Ji J. Investigation on the Current Situation and Countermeasure Research of Direct Drinking Water Hygiene Management in Schools of Qingdao City. Chin J Health Superv. 2016;23:283–6.

  17. Liu X, Chen H, Hu Y, Zhu Y, Meng T. The hygiene status of direct drinking water through pipelines in Haidian District, Beijing in 2016. J Appl Med. 2019;25:459–60+464.

  18. Ceng Q, Jiang Y, Li Y, Zou W, Zhang J, Yu B. Investigation and risk analysis of microbial contamination in the final rinse water of endoscopes in medical institutions. Chin J Disinfect. 2022;39:203–5+210.

    Google Scholar 

  19. Wu D, Zhao M, She T, Zhang K. Research on the application of the improved terminal pipeline of medical purified water in the endoscopy center in the quality management of purified water. Chin J Nosocomiol. 2020;30:1757–60.

    Google Scholar 

  20. Wang W, Ma J. The relevant reasons and countermeasures for the failure of cleaning and disinfection of digestive endoscopes. Chin J Nosocomiol. 2017;27:4077–80.

    Google Scholar 

  21. Silva CV, Magalhães VD, Pereira CR, Kawagoe JY, Ikura C, Ganc AJ. Pseudo-Outbreak of Pseudomonas aeruginosa and Serratia marcescens related to bronchoscopes. Infect Control Hosp Epidemiol. 2003;24:195–7.

    Article  PubMed  Google Scholar 

  22. Chroneou A, Zimmerman SK, Cook S, Willey S, Eyre-Kelly J, Zias N, et al. Molecular Typing ofMycobacterium chelonaeIsolates from a pseudo-outbreak involving an automated bronchoscope washer. Infect Control Hosp Epidemiol. 2008;29:1088–90.

    Article  PubMed  Google Scholar 

  23. Guimarães T, Chimara E, do Prado GVB, Ferrazoli L, Carvalho NGF, Simeão FC dos S, et al. Pseudooutbreak of rapidly growing mycobacteria due to Mycobacterium abscessus subsp bolletii in a digestive and respiratory endoscopy unit caused by the same clone as that of a countrywide outbreak. Am J Infect Control. 2016;44:e221-6.

    Article  PubMed  Google Scholar 

  24. Rosengarten D, Block C, Hidalgo-Grass C, Temper V, Gross I, Budin-Mizrahi A, et al. Cluster of pseudoinfections with Burkholderia cepacia associated with a contaminated washer-disinfector in a bronchoscopy unit. Infect Control Hosp Epidemiol. 2010;31:769–71.

    Article  PubMed  Google Scholar 

  25. Li Z, Li S, Li L, Yang Y, Chen W, Zhang Y, et al. Problems and countermeasures found during the acceptance process of the new endoscopy center. Chin J Infect Control. 2019;18:69–73.

    Google Scholar 

  26. Wang P, Han M, Shen G, Li N, Zhang X, Pan Y, et al. The disinfection effect of continuous hypochlorous acid disinfection on the final rinse water of endoscopes. Chin J Infect Control. 2022;21:617–22.

    CAS  Google Scholar 

  27. Zhang Y, Zhou Y, Cheng Y, Zhang X. Design of a sterile water preparation device for endoscope terminal rinsing. China Med Devices. 2020;35:57–9+69.

    CAS  Google Scholar 

  28. National Pharmacopoeia C. Pharmacopoeia of the People’s Republic of China (2020 Edition): part 1. Beijing: China Medical Science and Technology Press; 2020.

    Google Scholar 

  29. Walker JT, Bak A, Marsden G, Spencer W, Griffiths H, Stanton GA, et al. Final rinse water quality for flexible endoscopy to minimize the risk of post-endoscopic infection. Report from Healthcare Infection Society Working Party. J Hosp Infect. 2022;124:79–96.

    Article  CAS  PubMed  Google Scholar 

  30. Standardization Administration of China. Standards for Drinking Water Quality (GB5749-2006). Beijing; 2006.

  31. Zhang X, He Z, Wu Y, Lin Y, Zhao Z, Yuan J. Quality monitoring of medical water disinfection in medical institutions in Guangzhou City. Chin J Disinfect. 2022;39:581–3.

    Google Scholar 

  32. Wang W, Ma J. Investigation on the use of terminal rinsing water for endoscopes in 67 medical institutions and analysis of related factors. Chin J Nosocomiol. 2019;29:1587–90.

    Google Scholar 

  33. China Food and Drug Administration. Water for hemodialysis and related treatments (YY0572-2015). Beijing; 2015.

  34. Ji X, Ning P, Fei C, Zhang Y, Liu J, Liu H, et al. Research on the evaluation method of reprocessing quality of flexible endoscopes. Chin J Disinfect. 2021;38:552–4.

    Google Scholar 

  35. Wang X, Fan J, Shen Y, Wu X, Tan Z, Zhang Y, et al. Investigation on microbial contamination and disinfection effect of the centralized pure water supply system in a certain hospital. Chin J Disinfect. 2019;36:913–5.

    Google Scholar 

  36. Ning P, Ji X, Ding J, Fei C, Zhang L. Applied research on monitoring the disinfection effect of endoscopes by different detection methods. China J Endosc. 2019;25:67–71.

    Google Scholar 

Download references

Acknowledgements

Thank you to all the healthcare organisations that participated in the research.

Funding

This work was supported by Chinese Preventive Medicine Association Hospital Infection Department Development Youth Talent Promotion Project (CPMA-HAIsC-2024012900108); Jiangsu Provincial Association for Science and Technology Young Science and Technology Talent Support Project (Health Field)(JSTJ-2023-WJ006); Jiangsu Science and Technology Think Tank Program (Youth) Project (JSKX24055); Jiangsu Province Hospital Management Innovation Research Project (JSYGY-3–2023-559); Project of Chinese Hospital Reform and Development Institute, Nanjing University (NDYG2023039); The third Outstanding Young and Middle-aged Talents Training Program of Jiangsu Provincial People’s Hospital (YNRCQN0314); Young Scholars Fostering Fund of the First Affiliated Hospital of Nanjing Medical University (PY2022017).

Author information

Authors and Affiliations

Authors

Contributions

Z.L. and B.L. (as co-first authors) wrote the main manuscript text. Q.Z., Z.G., F.Z., W.C., and Y.Z. contributed to data analysis and interpretation. X.D. and J.D. prepared figures and tables. W.Z. supervised the project and revised the manuscript. All authors reviewed the manuscript.

Corresponding authors

Correspondence to Zhanjie Li, Jing Ding or Weihong Zhang.

Ethics declarations

Ethics approval and consent to participate

This study does not require ethical approval as it does not involve patient information or harm patients’rights. The research is limited to the analysis of endoscope rinse water and does not include human subjects or personal data.

Consent for publication

This study does not require informed consent as it does not involve human subjects, patient data, or any procedures that would necessitate consent. The research is limited to the analysis of endoscope rinse water and does not include personal information.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Z., Liu, B., Zhu, Q. et al. Management and microbial monitoring of final rinse water for flexible endoscopes in 290 hospitals in Jiangsu Province, China: a multicenter cross-sectional study. Antimicrob Resist Infect Control 14, 40 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13756-025-01560-3