Skip to main content

Table 1 Number of flexible endoscopes (FEs) without a working channel depending on the number of colony forming units (CFUs) before disinfection by disinfection group (p = 0.02)

From: A multicenter study comparing the bacterial reduction on flexible endoscopes without a working channel between UV-C light disinfection versus standard endoscope Washer Disinfection: a randomized controlled trial

Disinfection method

0 CFUs

1–50 CFUs

51–500 CFUs

> 500 CFUs

Total

UV-C light disinfection

(% within disinfection method)

61 (20.3)

163 (54.3)

62 (20.7)

14 (4.7)

300 (100)

Endoscope Washer Disinfector (% within disinfection method)

38 (16.5)

191 (59.0)

51 (18.8)

20 (5.7)

300 (100)